MAR Diagnostic Kit for Anti-Sperm Antibody
For IN VITRO Diagnostic use.


SPECIFICATION AND INTRODUCTION

[INTRODUCTION]
A kind of diagnostic kit to identify the infertile occurred
by anti-immune anti-sperm antibodies, on the basis of
the Mixed Anti-globulin Reaction (MAR) test, which is
quick and intuitionistic diagnostic kit.

[ACTION]
Mixed Anti-globulin Reaction (MAR) test is the standard method recommended by WHO for the detection of anti-sperm antibody (AsAb). The kit, based on classical Coombs Test, is to detect IgG-AsAb (In which case IgA almost does coexist with IgG) by the method of MAR test, and exerts actions of the invitro diagnosis of immunological infertility and the auxiliary diagnosis of seminalduct ligation complication.

Reference of Standard Inspection Items of Male Sperm (WHO)
Item Value
Quantity 2.0 ml or more
PH Not less than 7.2
Sperm Density 20X106 sperm/ml or more
Total Number of Sperm 20X106 sperm/l ejaculatio or more
Sperm Motility In 60 minutes after ejaculation, 50% or more sperm has movement of class a and class b or 25% or more has movement of class a
Modality ------
Fraction of Surviving 50% or more
Leucocyte Number Less than 1X106/ml
IBT Test Less than 50% sperm cling on the bead
MAR Test Less than 10% sperm cling on the bead
Comparison of MAR test and ELISA test for examining Anti-sperm Antibody
MAR ELISA ADVANTAGE OF MAR
Antigen Live sperm as antibody Memberane antibody of sperm Intuition
Examine antibody Antibody bond with live sperm Dissociation antibody in body fluid
Time About 10 minutes About 2 hours Simple and cheap
Instrument Microscope Microplate Reader
Usage For single sample For multi-samples Agility
Specificity High specificity Low Only sperm with antibody could bond to RBC, so easy to observe and test kinds of samples
Judgment According to the connection of sperm and sRBC According to the comparison of the color of reaction of sample, negative and positive
Sample Sperm, serum, cervical mucus and sperm plasma Serum, cervical mucus and sperm plasma
Shelf life 12 months, stability 6 months, less stability Reliable result based on stable quality
Comparison of MAR test and IBT test for examining Anti-sperm Antibody
MAR-AsAb IBT-AsAb P
Positive Rate 8.5% 8.1% P>0.05

[UNITS OF KIT]
a. Solution A 0.5 ml
b. Solution B 0.5ml
c. Solution C 2.5mlX2
d. Slide 10 pieces
e. Coverslip 20 pieces
f. One-off pipette 20 pieces
g. Rubber nipple 2 pieces
h. Instruction 1 sheet

[PROCEDURE]
A. Direct MAR Test
1. After being liquefied at the temperature of 37ºC for 30 minutes, 20 ul of fresh semen is put on slide and mixed with 20 ul of Solution A adequately for 15 seconds.
2. Add 20 ul of Solution B and mix adequately, put a coverlip on the slide. After 3 minutes, observe the results under the Optics Microscope (40 X 10) or Phase Contrast Microscope.
3. Recount after 10 minutes.

B. Indirect MAR Test
1. Centrifuge the sample with negative result in Direct MAR Test.
2. Wash the sample twice with Solution C, then balance the density of spermatozoa to 5X107 /ml.
3. 50 ul of the balanced sperm suspension is added with 200 ul of serum diluted by solution C (1:4), and incubate at the temperature of 37ºC for 1~1.5 hours. The mixture is centrifuged by Solution C in order to remove the dissociative Ig, then balance to 100 ul. The next process and judgment is the same as the Direct MAR Test.

[JUDGEMENT]
If without AsAb on the surface of the spermatozoa, the spermatozoa will swim freely among the adhered sRBC.
If AsAb exist on the surface of the spermatozoa, the sensitized will adhere to the spermatozoa. Under the condition of strong positive, the agglutinates become very massive and the spermatozoa will only twist in the agglutinates marked with sRBC.
At least 100 active spermatozoa should be counted, and calculate the percentage of the active spermatozoa with sRBC compared with the total active spermatozoa.
The Positive Ratio; R = (The quantity of active spermatozoa with sRBC / The quantity of total active spermatozoa) X 100%:
(1) R<10%, Negative
(2) R>10%, Positive
(3) R>40%, Strong Positive. Inferetility caused by AsAb should be considered especially.

[PRECAUTION]
a. Please read this Instruction carefully before using.
b. Fresh semen only.
c. Agitate the Solution A adequately before using.
d. All the solution should be balanced to room Temp., and the crystal in Solution C should be dissolved completely before using.
e. Pipette could be used to smear for twice in order to agitate adequately. The proper area is about 1~2 cm2.
f. Adverse influence on the test caused by the temperature should be noticed.

[STORAGE]
Stored at 2~8ºC

[EXPIRY]
12 months 

[LICENSE NUMBER]
GYZZS20030032

[MANUFACTURED BY]
Anhui Anke Biotechnology (Group) Co., Ltd.

[ADDRESS]
Ankebio Building, 669 Changjiang Road, West
Hefei, 230088, P.R. China.
tel: +86-551-5319890, 5318811
fax:+86-551-5319895
e-mail:ankebio@mail.hf.ah.cn
http://www.ankebio.com

 
 
Copyrights 2006 Ankebio Co.,Ltd. All rights reserved.  Disclaimer Licenced Number ICP 05016497.     Design by:IBW
  Tel:0086-551-5319890    Fax:0086-551-5319895
E-mail:ankebio@mail.hf.ah.cn